GA50 VOCs FID Gas Analyzer | HighSeek
Product Overview
The GA50 Online FID Gas Chromatograph is designed for continuous monitoring of VOCs in ambient air and industrial gas streams.
Building on established gas chromatographic technology for total hydrocarbon and non-methane hydrocarbon analysis, the GA50 integrates additional BTEX measurement capabilities to meet the growing demand for comprehensive VOC monitoring and air quality assessment.
The analyzer combines advanced gas chromatographic separation, multi-column configuration, column switching and backflushing technology, and high-sensitivity Flame Ionization Detectors (FIDs). This integrated analytical system enables the selective measurement of methane, total hydrocarbons, non-methane hydrocarbons, and individual benzene series compounds.
The complete sample path is maintained at a controlled temperature of 180 °C to minimize condensation, adsorption, and sample loss during transportation and analysis. With automated sampling, intelligent control, and real-time data processing, the GA50 provides reliable online monitoring for environmental protection, industrial emission monitoring, laboratory analysis, and process control applications.
Designed for continuous operation and flexible system integration, the GA50 supports remote monitoring, data transmission, and connection to plant-level or environmental monitoring platforms.
Working Principle
Gas Chromatographic Separation with Dual FID Detection
The GA50 employs gas chromatography (GC) combined with column switching, backflushing, and dual Flame Ionization Detectors (FIDs) to measure methane, non-methane hydrocarbons, and BTEX compounds.
The sample is continuously transported through a heated sampling system maintained at approximately 180 °C. After filtration, a precisely controlled sample volume is introduced into the chromatographic system through an automated sampling valve and quantitative sampling loop.
Driven by the carrier gas, the sample passes through specially configured chromatographic columns, where target compounds are separated according to their chromatographic retention characteristics.
The analytical process consists of three main measurement paths:
1. Total Hydrocarbon Measurement
The first analytical path directs the hydrocarbon components to an FID detector for total hydrocarbon measurement.
2. Methane and Non-Methane Hydrocarbon Measurement
The second path uses chromatographic separation and backflushing technology to separate methane from other hydrocarbon components. Methane is measured by the FID, while non-methane hydrocarbon concentration is determined from the corresponding total hydrocarbon and methane measurements.
3. BTEX Measurement
The third path uses a pre-column and separation column configuration to separate benzene, toluene, ethylbenzene, and xylene compounds. The separated components are then measured by the FID detector. Unwanted heavy hydrocarbon components are removed through the backflushing process.
Through these integrated analytical paths, the GA50 provides automated and continuous measurement of methane, NMHC, and individual BTEX compounds.
Key Features and Advantages
1. High-Sensitivity Dual FID Detection
Equipped with dual high-sensitivity Flame Ionization Detectors, the GA50 provides reliable detection of low-concentration hydrocarbon compounds.
The system incorporates automatic startup diagnostics, power-loss protection, automatic restart, automatic ignition, and automatic calibration functions to support stable and repeatable operation during continuous monitoring.
2. Advanced Multi-Column Configuration
A specialized chromatographic column combination and switching design enables efficient separation of methane, non-methane hydrocarbons, and BTEX compounds.
The modular analytical architecture and highly integrated electronic control system improve operational stability, simplify system integration, and facilitate maintenance.
3. Automated Sampling and Intelligent Data Processing
The GA50 supports fully automated online sampling and analysis using a dual 10-port valve configuration.
Its integrated control and data processing platform provides:
- Real-time monitoring of analytical data and chromatograms
- Automated sampling and measurement sequences
- Remote instrument operation and control
- Real-time data transmission
- Historical data recording and analysis
- Integration with external data acquisition and supervisory control systems
Flexible sample introduction interfaces allow the analyzer to be integrated into different online sampling configurations.
4. High-Temperature Sample Handling
The complete sample transfer path is maintained at approximately 180 °C to minimize condensation, adsorption, and sample residue.
This heated sampling design helps preserve sample integrity, particularly when measuring trace-level VOCs and hydrocarbon compounds.
5. Fast Analysis and Wide Dynamic Range
Optimized chromatographic separation and column switching technology enable rapid analysis of NMHC and BTEX components.
The analyzer offers a wide linear measurement range of up to 10⁷, supporting applications involving both low-concentration environmental monitoring and higher-concentration industrial gas analysis.
6. Flexible System Integration
Multiple communication and signal interfaces are available for integration with environmental monitoring systems, industrial control platforms, and centralized data management systems.
Supported interfaces include 4–20 mA analog output, RS232, RS485, and Ethernet.
Technical Specifications
| Parameter | Specification |
|---|---|
| Product Model | GA50 |
| Product Type | Online Gas Chromatograph |
| Measurement Principle | Gas Chromatography (GC) with FID Detection |
| Detection Technology | Dual Flame Ionization Detectors (FID) |
| Target Compounds | Methane, Non-Methane Hydrocarbons (NMHC), Benzene, Toluene, Ethylbenzene, Xylenes |
| Methane Measurement Range | 0.1–10,000 ppm (v/v) |
| NMHC Measurement Range | 0.04–10,000 ppm (v/v) |
| Benzene Measurement Range | 0.1–10,000 ppm |
| Methane Detection Limit | ≤0.1 ppm (v/v) |
| NMHC Detection Limit | ≤0.04 ppm (v/v) |
| Benzene Detection Limit | ≤0.1 ppm |
| Measurement Repeatability | RSD ≤3% |
| NMHC Analysis Time | <1 min |
| BTEX Analysis Time | <3 min |
| Sample Path Temperature | Approximately 180 °C |
| Linear Dynamic Range | Up to 10⁷ |
| Analog Output | 4–20 mA |
| Digital Communication | RS232, RS485, Ethernet |
| Power Supply | 220 VAC, 50 Hz |
| Power Consumption | <600 W |
| Operating Temperature | 5–50 °C |
| Operating Humidity | 10–90% RH |
Typical Applications
1. Ambient Air Quality and Environmental Monitoring
- Continuous monitoring of methane and non-methane hydrocarbons in ambient air
- BTEX monitoring for environmental air quality assessment
- VOC monitoring in urban and industrial environments
- Air pollution monitoring stations
- Environmental compliance monitoring and pollution source assessment
2. Industrial Emission and VOC Monitoring
- VOC emission monitoring in industrial facilities
- Hydrocarbon emission monitoring in petrochemical plants
- BTEX monitoring around chemical production facilities
- Fugitive emission assessment and pollution source identification
- Industrial exhaust gas monitoring and environmental management
3. Petrochemical and Refining Industries
- Hydrocarbon composition monitoring in refinery processes
- Methane and NMHC analysis in petrochemical production
- Benzene series monitoring in chemical processing facilities
- Continuous monitoring of hydrocarbon-containing gas streams
- Process gas composition analysis and operational optimization
4. Laboratory Analysis and Research
- Laboratory analysis of methane, NMHC, and BTEX
- Gas chromatographic analysis of hydrocarbon samples
- Environmental analytical research
- Method development and analytical performance evaluation
- Reference measurements and instrument comparison
5. Industrial Process Monitoring and Control
- Continuous hydrocarbon composition monitoring in production processes
- VOC concentration monitoring in industrial gas streams
- Process gas quality inspection
- Hydrocarbon process optimization and operational control
- Integration with distributed control systems (DCS) and supervisory monitoring platforms
Why Choose the GA50 Online Gas Chromatograph?
The GA50 combines gas chromatographic separation, dual FID detection, automated sampling, and intelligent data processing in one integrated online analytical platform.
Its ability to measure methane, non-methane hydrocarbons, and individual BTEX compounds makes it suitable for comprehensive hydrocarbon and VOC monitoring across environmental and industrial applications.
With heated sample handling, automated column switching, fast analytical cycles, and flexible communication interfaces, the GA50 provides a practical solution for continuous VOC monitoring, industrial emission assessment, and hydrocarbon process analysis.